Troubleshooting#

Common problems and their solutions, organized by what you see.


Connection Issues#

Device not detected after connecting USB#

  1. Check that both the power cable and USB cable are securely connected.
  2. Try a different USB port on your computer (preferably USB 2.0 or higher, directly on the machine — not through a hub).
  3. Restart the NMR device, wait 10 seconds, then restart TQT Nuclei.
  4. On Windows, open Device Manager and check for unknown devices under "Ports (COM & LPT)". If the device appears with a warning icon, reinstall the USB driver.
  5. On macOS, check System Information → USB to verify the device is recognized by the OS.

"Connect" button does nothing#

Make sure Fake Mode is turned off in the HOME page connection panel. Fake Mode simulates a device and disables real hardware connections.


Temperature Issues#

Red temperature indicator#

A red signal means either overheating or a sensor error:

  1. Power off the device immediately.
  2. Check that the device is in a well-ventilated area, away from direct sunlight or heat sources.
  3. Wait 15 minutes, then power on again.
  4. If the red indicator persists, the temperature sensor may need service — contact support.

Temperature won't stabilize (stays yellow)#

  • Normal warm-up takes 1-2 hours. In cold environments (below 15°C), it can take longer.
  • Make sure the device is not near air conditioning vents or open windows — drafts cause temperature fluctuations.
  • If T_cur oscillates around T_set without settling, the PID controller may need adjustment — contact support.

Signal Quality#

No signal at all#

  1. Check frequency tuning. An incorrect frequency is the most common cause. Run HOME → FREQUENCY auto-tune or press AUTO TUNE.
  2. Check the sample. Make sure the sample tube is properly inserted and contains enough material (at least 1 cm of liquid above the coil center).
  3. Check pulse durations. If p90 is wildly wrong, you may get zero signal. Re-run HOME → DURATION tuning.
  4. Check the receiver gain. If RX Gain is too low, the signal may be below the noise floor. Try increasing it in SETTINGS → Device.

Weak or noisy signal#

  • Increase the number of scans — signal-to-noise ratio improves with the square root of the number of scans. Doubling scans gives ~40% better SNR.
  • Enable phase cycling — this cancels certain systematic artifacts. Set Num Scans to a multiple of 4.
  • Check dead time — if it's too short, hardware ringing contaminates the signal. Re-tune in HOME → DEAD TIME.
  • Check temperature stability — measurements during warm-up (yellow indicator) will have poor reproducibility.

Baseline drift or offset#

  • Turn on Enable auto baseline correction in SETTINGS → Preprocessing → Baseline Correction.
  • Check that the phase is properly tuned — a misaligned phase creates a sloping baseline. Re-tune in HOME → PHASE.
  • If the drift is periodic, check for nearby sources of electromagnetic interference (motors, transformers, fluorescent lights).

Experiment Issues#

T1 or T2 values seem wrong#

  • Compare with literature values for your sample. Water at 25°C: T1 ≈ 2.5 s, T2 ≈ 2.5 s. Dissolved oxygen, impurities and temperature all shift these values, so a large difference is a prompt to check the measurement, not proof of an error.
  • Check the repetition delay — it should be at least 5× T1. If it's too short, T1 will appear shorter than it really is.
  • Check the echo time (for T2) — a very long echo time can cause signal loss before the first echo, biasing T2 downward.
  • Verify the fit — check the R² value. A low value suggests the fitting model may not match the data (try multi-exponential instead of single-exponential).

CPMG echoes not forming properly#

  • Re-tune p180 pulse duration — an inaccurate 180° pulse produces incomplete refocusing.
  • Check echo time — if τ is too short, pulses may overlap with the dead time.
  • Reduce decimation to capture more echoes and diagnose where the train breaks down.

Custom sequence produces no output#

  • Make sure your sequence ends with an ADC() node/call — without it, no data is acquired.
  • Check the Sequence preview graph at the top of the page to verify the timing looks correct.
  • Ensure seq_null is set correctly — it should be the total time before the first ADC window.

Software Issues#

License file not accepted#

  • Verify the file has a .tl extension and hasn't been renamed or corrupted during download.
  • Check that your license hasn't expired — trial licenses are valid for 14 days.
  • If you recently reinstalled the OS, the machine ID may have changed. Contact support for a license reset.

A known problem. A paid plugin's licence is tied to a fingerprint of this computer, and part of that fingerprint is the network hardware address and the computer name. Both can change without anything being wrong:

  • macOS Private Wi-Fi Address (and similar settings on Windows and Linux) rotates the Wi-Fi address.
  • Plugging in or removing a dock, a USB network adapter or a virtual-machine network changes which address is used.
  • The computer name can change with the network you are on.

Updating Nuclei restarts it, so this often shows up right after an update even though the update itself changed nothing.

To fix it, open Plugins, uninstall the plugin and install it again from the marketplace while signed in. This issues a licence for the current fingerprint. If you see "Machine limit reached", contact support to free the old machine entries.

Application crashes on startup#

  • Restart the application. Do not delete the application data folder to fix a crash: it can hold your profiles and settings. Contact support first.
  • Update to the latest version from terraquant.tech.
  • If the crash persists, check the log file (accessible from the status bar at the bottom of the main window) and include it when contacting support.

Slow performance or UI lag#

  • Close other resource-heavy applications.
  • Large experiment files (>100 MB) can slow down the Data Browser — consider archiving old projects.
  • Plots draw every point; nothing is thinned. Spectra of about a million points and large 2D planes take longer to display, to phase automatically and to peak-pick, especially with software rendering.
  • On macOS, ensure TQT Nuclei has been granted necessary permissions in System Preferences → Security & Privacy.

Still stuck?#