DOSY#
The DOSY page turns a diffusion-ordered dataset into a map of chemical shift against diffusion coefficient, so that the resonances of each molecule in a mixture line up at their own diffusion rate.
Open it from the navigation sidebar, or from the DOSY button in the spectrum tool sidebar while a diffusion dataset is on screen. That button only appears for a pseudo-2D scan whose experiment type is DOSY — the indirect axis has to carry gradient strength for there to be anything to invert.

How a run works#
- Load a diffusion scan from the Data Browser.
- Check the acquisition parameters in the left rail — the gradient ramp, δ, Δ and γ. They are read from the acquisition metadata, but a value taken from the wrong slot in the pulse programme scales every diffusion coefficient by a constant while leaving every fit looking perfect. Correct anything that is wrong before processing; corrected fields are marked as yours rather than the instrument's.
- Choose an inversion method and press Process.
- Read the map, then inspect individual columns or integral regions.
Every diffusion coefficient is derived from b = (γ·g·δ)²·(Δ − δ/3). If the resulting b range falls far outside 10⁸–10¹¹ s/m², the page says so — a standard probe does not reach beyond that, and the numbers should not be trusted until δ and Δ are checked.
Inversion settings#
| Setting | What it does |
|---|---|
| Method | Regularised NNLS (CONTIN) and Maximum entropy recover a distribution of diffusion coefficients per column. Single exponential and Two exponentials are direct Stejskal–Tanner least-squares fits of one or two decays to each column, with no smoothing: they report each D, its share of the column, a one-sigma uncertainty from the fit, and the residual. A two-exponential fit needs at least five gradient increments, and its uncertainties grow large when the two components are close in D. |
| Regularisation | How strongly the recovered distribution is smoothed. Automatic (L-curve) picks it from the data. |
| Diffusion grid | Minimum D, maximum D and the number of grid points the distribution is recovered on. |
A column that cannot be inverted is left without a result rather than given a number. The quality strip reports why: too noisy, too little decay across the ramp, too few gradient increments, or below the intensity threshold. If nothing inverts at all, check the gradient ramp and the signal level first.
Reading the result#
The map is the page's main view: chemical shift horizontally, diffusion coefficient vertically, usually on a logarithmic D axis. Resonances belonging to one molecule share a horizontal band.
Two tabs sit beside it, Regions and Components, and below them the detail for the column you select in the map:
- Regions — the integral regions and their fits (see below).
- Components — the resolved components as a table: shift, D, weight, distribution width in decades, and hydrodynamic radius. After a direct fit, D carries its ± uncertainty.
- Decay (detail view) — the measured attenuation of one chemical-shift column against b, using the real gradient-derived b values, with the fit and the residual.
- Distribution (detail view) — the recovered P(D) for that column.
Changing a setting after a run marks Recompute in amber: the map on screen no longer matches the settings. Cancel beside the run button stops a running inversion between columns and leaves the map on screen as it was. Preview with synthetic data on the empty page shows a banner with Exit preview, and the preview leaves by itself when a real result arrives.

Integral regions#
A single column has a low signal-to-noise ratio; an integral over a resonance has a much higher one. Right-click and drag across the chemical-shift axis of the map to draw a region, exactly as you would integrate a 1D spectrum. Each region gets its own three-parameter decay fit (a baseline plus one exponential) and its own diffusion coefficient. With Two exponentials selected, regions instead get a two-component fit, and the table shows both diffusion coefficients with the slow component's share in the tooltip.
Suggest picks integral windows automatically from the least-attenuated increment, keeping any region you drew or renamed yourself. Suggested regions are badged as such until you edit them.
The regions table flags two problems worth acting on. A large baseline share means the window is not clean and its D is less reliable than the residual suggests. An overlap badge means a shared peak is being integrated into two regions at once.

Calibrating D to a reference#
The Calibration section of the Inversion tab scales every fitted D after the fit. Give the ppm window of a reference signal and either choose Water (HDO in H2O), whose self-diffusion coefficient is taken at the map's temperature from the fit of Holz, Heil and Sacco (Phys. Chem. Chem. Phys. 2000, 2, 4740; 0 to 100 °C), or choose Known D and enter the reference's own value. The window is fitted like a region; each D, the D axis and the hydrodynamic radii are multiplied by known D divided by measured D. The map's axis names the reference, the factor is recorded in the report data, and the exported map carries the same axis name and the factor in its title. Only compounds with a published value are tabulated; other liquids need their known D entered.
On the map, each region fit with a standard error is drawn as a diamond at the middle of its window with a bar of one standard error (a bi-exponential fit gives two).
Exporting the tables#
The download button beside the Regions and Components tabs writes the open table as CSV. Components lists each shift's diffusion coefficient, uncertainty, weight, fraction of the column and radius; regions lists each window's ppm range, diffusion coefficients, fractions, baseline and residual. Values are unrounded.
The map itself exports from the image button beside it, as a 2D file in the formats that carry a 2D spectrum (JCAMP-DX and a CSV matrix). Its F1 axis is log10 of D in m²/s, stated in the file, and its F2 axis is ppm; the intensities are those drawn on the map from the last run.
Hydrodynamic radii#
Given a temperature and a solvent viscosity, the page converts each diffusion coefficient to a hydrodynamic radius through the Stokes–Einstein relation, which assumes a rigid sphere with stick boundary conditions.
Temperature comes from the acquisition metadata when the instrument recorded it; viscosity comes from a table of common deuterated solvents. Either can be overridden. If neither a temperature nor a viscosity is available, radii are not estimated and the panel says so instead of guessing.
Note
The bundled sample dataset includes a DOSY scan of an ethylene glycol and caffeine mixture, which is enough to work through this page without an instrument. See Explore sample data.