Settings#

Settings holds the instrument parameters, the processing applied to data as it is captured, where data is stored, and how the app itself looks and behaves. It is organised into tabs along the top.

The Settings page, Device tab

Device#

The operating parameters of the connected instrument. The fields shown depend on the device: an unrecognised device falls back to a basic set, and with nothing connected the tab says so instead of offering controls that would go nowhere.

Group Fields
Sensor Frequency, phase, TX-1 gain.
Timing Repetition delay, dummy shots, phase cycle.
Receiver RX gain (2–8% is the recommended range), RX phase.
Pulse durations The duration of the 90°, 180°, 270° and 360° pulses, in microseconds.
Analog processing FIR window and type (Gauss or Sync), bypass FIR, low-pass filter cut-off, analog and digital IF, base level, dwell time.

Most of these are found for you on the Home page — frequency and phase in particular. The repetition delay is derived from the sample's anticipated T1 and recalculated whenever that changes.

Warning

The IF frequencies and the dwell time are set by the manufacturer. Dwell time is fixed; changing the IF values without reason will stop the instrument detecting signal. Reset Settings lists the fields that differ from their defaults and, once you confirm, restores them.

Every field carries a tooltip with its range and its effect. Read those before changing anything on this tab.

Preprocessing#

What is applied automatically to a signal after capture, before you see it.

  • Zero filling — on or off, and the factor. A factor of 0 pads to the next power of two; otherwise it multiplies the length. A factor below 1 does nothing, since the length is never reduced.
  • Line-shape correction — off by default; 1D FIDs only. Reference deconvolution: one clean reference line is isolated, divided out of the FID and replaced by an ideal line (Lorentzian or Gaussian) of the target width, so every line loses the lineshape distortion and drift the instrument added. It runs on the FID before linear prediction and apodization. The fields are the reference region (ppm, on the axis as acquired; empty takes the sample's internal standard, else its solvent line, found in the signal), the target line width in Hz (empty keeps the width the reference line measures, which corrects its shape without narrowing it) and the ideal shape. In the plot's Processing panel the region can also be drawn: choose Draw on the spectrum, then drag with the right mouse button across the reference line (Esc cancels). The reference line must stand well above the noise and nothing else may lie in the region, otherwise the stage is skipped with a message. Line areas are preserved only as far as the one reference line is clean, so qNMR warns when the stage is applied. It does not apply to 2D data or to data that has no FID.
  • Linear prediction — off by default. Extends the FID forward by predicted points, or replaces its first points (backward), before apodization. The fields are the direction, the number of coefficients (default 32), the points to append or replace, the number of leading points the coefficients are fitted on (default all), and, for 2D data, the axis (F2, F1 or both). A forward extension is limited to 8 times the acquired length. The stored FID is not changed. The stage does not apply to data that has no FID, and F1 is skipped for parameter axes and echo-antiecho data. A scan can override it in the plot's Processing panel.
  • Apodization — the window function (exponential, Gaussian, Hamming, Hann, Blackman, Blackman-Harris, Kaiser, Bartlett, cosine, Tukey, sine bell, trapezoid, TRAF and Lorentz-to-Gauss) and its width. Line broadening is a Lorentzian FWHM in Hz; adaptive line broadening estimates it from the natural line width in the data instead. The sine bell takes a start (a fraction of pi; 0.5 is a cosine) and a power (1 for a sine bell, 2 for a squared sine bell); the trapezoid takes the fraction of the FID it rises over and the fraction it falls over; TRAF (Traficante and Nemeth) takes the Hz line broadening of an exponential and shapes it by its mirror image, so it stays near 1 for most of the FID and rolls off at the end; Lorentz-to-Gauss enhances resolution, with line narrowing in Hz and a Gaussian width in Hz, at the cost of noise. Every window is stored with the rest of the apodization settings, so it can be saved and reapplied by name; Resolution enhancement, Sensitivity enhancement and Squared sine bell are built in and read-only.
  • FFT — the transform method.

Note

qNMR requires no line broadening. Set it to 0 before integrating quantitatively.

Anything beyond this is better done in the Processor, where the treatment is explicit and reusable.

Databases#

The compound databases consulted for peak identification and by structure elucidation. The built-in database is always available; add PostgreSQL connections or JSON files for broader coverage.

My spectra#

My spectra is a source over the user's own projects, listed and enabled by default beside the other sources. It stores no spectra: it is an index of scans that already exist in projects, kept in the data root in the Spectral Index folder. Deleting the index loses nothing; it is rebuilt from the projects.

For each indexed scan the index holds the reported shifts of a 1D spectrum (with pattern, couplings and integral), the structure of the sample with its substructure fingerprint, the name and formula, and where the scan is (project, sample, scan id).

Option Effect
Scope By default only scans marked as references are searched. All projects searches every indexed scan that has peaks or a structure.

The index updates when a scan is written (import, processing, peak or structure edits) and when a project is opened or saved. Updates run one scan at a time in the background. A result whose project or scan can no longer be found is shown as unavailable; Rebuild index reads every project again.

Every place that queries databases (the NMR Database page, compound identification, structure elucidation) uses this source when it is enabled, and a result names its source. A result opens its scan, or overlays it on the current plot. Scores depend only on how well the peaks or structure match, never on how common a structure or shift is.

Searches on this source:

Search Query
Peak list Shifts with a ppm tolerance (default 0.05 for 1H, 2 for 13C); scored by the shared peak matcher.
Structure A SMILES typed in the page search finds the scans whose structure contains it.
Name or formula Part of the scan or sample name, or an exact Hill formula.

Sharing references uses the .nuclei export of the marked scans.

Weight rare shifts higher (off by default) applies to every source: when on, matching gives more weight to shifts that few compounds share; when off, no frequency weighting is used.

Each connection can be tested — the test reports the compound and peak counts it can see — and enabled or disabled without removing it. Removing a connection stops lookups that relied on it from returning results; the database itself is untouched.

Python packages#

Install Python packages from PyPI for use in analysis scripts, plugins and processor code snippets. Type a package name and press Install, or search and install from the results. Packages you install are available to all of these. A failed install shows the reason next to the package.

Uninstalling a package asks for confirmation, and warns you which plugins depend on it. Scripts and plugins that import a removed package stop working until it is installed again.

Appearance#

Setting What it does
Navigation bar on the left Shows the page list as a bar beside the content, in addition to the top bar. Off by default. While it is on, the top bar drops its page shortcuts and the page menu, and the bar lists the pinned pages first, in pin order, above the rest. ⌘1–⌘9 jump to the pinned pages in either layout. Right-click an entry to pin or unpin it, or drag a pinned entry to reorder the pins. The page list, its order and hidden entries are the same in both.
Colour theme System, Dark, Light, Warm Paper, High Contrast Dark or High Contrast Light, with separate choices for the light and dark slots. See Accessibility.
Language The interface language. New installs follow the operating-system language.
Start page Which page opens at launch.
Frequency display unit Hz, kHz, MHz or GHz.
UI scale Zoom level for the whole interface.
Splash screen Whether the start-up sound is muted. Press any key or click to skip the splash.
Reduce animations Follow system, Always or Never. Disables transitions, dialog animations, backdrop blur and animated graph edges.
Teaching features Admin only, off by default. Shows the teaching tools; see Teaching. Can be locked by your organisation.
Popout panels Whether popped-out panels stay above the main window, and whether they reopen at the next launch.

Pinned pages at first start. The first-start setup asks which paths apply (run an experiment, make sense of a spectrum, quantify, automate and QC, work out a structure, review, measure T1, T2 and series, and teaching when teaching features are on). The paths set the pinned pages: each path lists its pages by importance, the lists are merged by taking every path's first page, then every second page, and so on, duplicates are dropped, at most five are kept, and they are shown in work order (device, sample, sequence, acquisition, processor, analysis, reports, reviews). Elucidation is pinned only where the edition includes it, and Connect or Learn only when teaching features are on. With no path chosen the pins are Acquisition, Data Browser, Processor, Analysis and Reports. Pins changed afterwards are not overwritten when the setup is run again.

Customize navigation… reorders the page list and hides entries you do not use, in the top bar's page menu and in the left navigation bar alike. Settings and Help cannot be hidden. Entries for pages that are no longer installed are listed separately so you can clear them out.

The Appearance tab

Explain mode#

Explain processing steps adds a What happened and why box to the Processing History panel: a plain-language note on what each processing step did to the scan, with links into the glossary. It is off by default and available to every profile. See Accessibility.

Storage#

Where NucleI keeps profiles, projects and scans. Data folder in use shows the current folder. Every profile sees this tab; only an Admin can change it, and the others see Only an administrator can change the data folder.

To move the data folder:

  1. Type a full path in Data folder, or use Browse…. Use default goes back to the standard location.
  2. Press Apply.
  3. If the current folder holds data, Move the existing data? asks what to do with it: Move data moves it to the new folder at the restart, Leave data starts with the new folder as it is, and Cancel changes nothing.
  4. Enter the Administrator password in Confirm with the administrator password.

The change applies at the next start. Until then a banner says what will happen, with Restart now and Cancel change. A move is refused if the new folder already contains files or if one folder is inside the other; the data folder then stays where it was, and the banner reports why at the next start.

Note

If your organisation sets the data folder by machine policy, the field is locked and shows Set by your organisation. The NUCLEI_DATA_ROOT environment variable also overrides this setting, and a banner says so. See IT deployment.

Reports#

A signature stamp — a picture of your handwritten signature — placed in the signoff section of reports generated from this profile. Accepts PNG, JPEG, TIFF, BMP or WebP up to 8 MB; the image is re-encoded on storage, which strips camera and location metadata.

Warning

A signature stamp is a picture, not a cryptographic signature. It proves nothing on its own, anyone with access to this profile can put it on a report, and it is embedded in every report exported from this machine.

Current view controls whether reloading a sample re-captures the acquisition plot into the report's signal and spectrum placeholders. It is off by default, because turning it on replaces figure options you set by hand.

Telemetry and privacy#

Four independent choices: anonymous crash reports and usage statistics; detailed diagnostics, which add system information and a pseudonymous device identifier; contributing extracted peak data with the sample name, experiment type and instrument frequency; and contributing full processed spectra. Anonymous crash reports and usage statistics are on by default; you can change this in the first-run setup or here at any time. The other three are off unless you turn them on. No file contents or personal data are sent.