qNMR#
The qNMR page turns integrals from a 1D spectrum into a purity or concentration with an uncertainty budget and a suitability record. It is a stepped workflow: the step list on the left shows what is done and what is still missing, the centre pane holds the current step, and the right pane shows the result as soon as there is one.
Opening it#
Open the qNMR button in the spectrum view tool sidebar while a 1D spectrum is shown, or the qNMR entry in the navigation bar. The regions you quantify are the integration regions drawn on the current scan, so draw them in the viewer first; the page itself does not draw them.
The strip above the steps#
The strip at the top holds the method and the Run button.
| Method | Use it when |
|---|---|
| Internal standard | The standard is weighed into the same tube. |
| External standard | The standard is in a separate tube. |
| PULCON | One certified external reference, corrected for pulse length and temperature. |
| Electronic reference | A calibrated instrument response replaces a reference compound. |
| 100 % method | Relative purity from the share of every quantified signal; no standard, no weighing. |
Hover a method for its hint. The strip also shows the suitability status and a Spectrum changed chip when the spectrum has been reprocessed since the last run. Run stays disabled while inputs are missing; the strip and the step list say how many (for example "2 inputs still needed") and name them.
The steps#
Click any step in the list to go to it. Each step shows a tick when complete, a warning when it still needs input, and an empty circle when not yet reached.
| Step | What you do or get |
|---|---|
| 1. Dataset | Read-only summary of the project, sample, scan and structure, plus the acquisition on record: Relaxation delay, Acquisition time, Transients, Dummy scans, Flip angle and Measured T1. A parameter the scan does not record is shown as Not recorded and its check is reported as unverified, not as acceptable. |
| 2. Processing | Propose a recipe asks for the line broadening, zero filling and baseline correction this spectrum should carry before integration. If the project's processing differs, the differences are listed, with a button to Preprocessing settings. |
| 3. Regions | Give each integration region a role: Analyte, Reference or Excluded. Draw regions in the viewer returns to the plot. Methods that use a separate reference scan also ask you to choose it (the scan under test cannot be its own reference). |
| 4. Suitability | Assess checks the acquisition (relaxation delay against T1, signal-to-noise, digital resolution, acquisition time, receiver headroom, steady state, excitation geometry) and each region (SNR, baseline, overlap, lineshape). Each check is Pass, Caution, Fail or Unverified. Opening this step also runs the assessment. |
| 5. Inputs | The numbers the method needs: names, molar masses (with Derive from the structure when the sample has a structure), nuclei counts, weighed masses, certified purity or concentration, response factor, solution volume, Replicate purities and the Coverage factor. Which fields appear depends on the method. |
| 6. Uncertainty | After a run: the budget table with each contributor's relative and absolute value and its Share of variance, plus a replicate summary when replicates were entered. |
| 7. Audit record | The settings and results of the run as a table of fixed keys. Copy as CSV or Export as CSV. |
| 8. Report | Add to report appends a qNMR section to the current report template; Open the Report Builder goes to the Report Builder. Audit record (CSV) and Uncertainty budget (CSV) export the raw tables. |
A successful Run moves you to Inputs; the result appears in the right-hand Result panel with the purity and its expanded uncertainty at the chosen coverage factor, and, where the inputs allow, concentration and mole fraction.
Suitability and caveats#
If a suitability check fails, the purity is shown as Purity (provisional) with the failing checks named, and the Report step asks you to accept the caveat first. Accept caveat records it; the result is then marked Caveats accepted and the audit record lists it. Review jumps back to Suitability.
If the spectrum is reprocessed after a run, the result is flagged Spectrum changed; run again before reporting it.
Advanced limits#
Under Advanced in the step list, the suitability limits (for example the recycle time target in multiples of T1, or the signal-to-noise target) and the transfer uncertainty between acquisitions can be changed. The defaults are the textbook figures; changed values are marked customised, recorded with the result, and Restore the defaults resets them.
Note
The result is stamped with the spectrum it was computed from. Re-phasing or reprocessing afterwards makes it stale rather than silently changing it.