Skip to content

Experiment Queue#

The Experiment Queue is a list view of a schedule: experiments run back to back, one row each. It lives on the Task Manager page and edits the same saved schedule as the node canvas, so there is no separate queue file or queue history.

Use the Queue / Graph switch at the top left of the page to move between the two views. The choice is remembered. Use the queue for a plain sequence of measurements, and the graph for branching, conditions or several instruments.

Note

Runs of a queue are recorded in Run History like any other schedule run. The queue does not keep its own log.

Rows#

Each row is one experiment and gives one record. The columns are Experiment, Project, Key parameters, Processing, Analysis and Status.

Add What it does
Add experiment Lists the instrument's own experiments (grouped as 1D, 2D, Relaxation, Setup, Other) and your Saved experiments. Instrument experiments need a connected instrument.
Wait Pauses for a duration above zero.
Repeat block Repeats the rows inside it a set number of times, or until stopped, with an optional pause between passes. An empty block cannot run.
Other step Adds an Analysis, Open project or Notify step. Other steps, such as Export results, are added in the graph and then show here as a row.

Drag a row to reorder it. The row menu offers Edit parameters, Duplicate, Delete (with undo), Run from here, and moving a row into or out of a block. Open project sets the sample the following rows are saved to, and Copy to next sample… copies a sample's rows to another project. Undo and Redo are in the toolbar.

Toolbar#

Control Meaning
Instrument The connected instrument the queue runs on. Click to connect or change it.
Run In the app: the app starts each experiment. On the instrument: the instrument runs the experiments from its own queue and can move autosampler tubes (a Tube choice appears on each row). Only instrument experiments can run there, not waits, repeats or saved experiments.
On error Stop, Skip and continue, or Stop and disarm. A lost instrument always stops the queue.
Lock before each experiment Engages the deuterium lock before every experiment.
Shim when the sample changes Shims before the first experiment of each sample (project). When the instrument runs the queue, its own settings decide.
Schedule… Runs the queue on its own: Once, Every interval or Weekly. The queue must be saved with a name. Save and arm, Disarm and Remove schedule work as for a trigger in the Task Manager.
Save as template Saves a copy with every experiment and setting but no projects, tubes or schedule, to reuse for other samples.

The toolbar also shows a duration estimate taken from the instrument's own estimates plus waits and repeats, and the number of problems. Rows without an estimate are counted separately.

Running#

Start the queue with the page's run button, or from one row with Run from here. Before it starts, problems are listed in This queue cannot run yet, for example a deleted saved experiment, an empty block, or an experiment made on a different instrument. The instrument runs one experiment at a time, so the queue will not start while another schedule is running on it.

While it runs, a row's status is Queued, Running, Done, Failed, Skipped or Stopped. Rows not yet started can still be added, changed, moved or removed; running and finished rows are locked. When the instrument runs the queue itself, edits no longer reach the run.

  • Stop after this step lets the running experiment finish, then stops.
  • Stop now halts the instrument immediately.
  • After a stop, Resume from row N continues at the next row. After a failure, Retry from row N reruns the failed row.

When the list is read-only#